In vivo dosimetry of stereotactic radiation therapy
Author:
Jianguo Qian, et al, Johns Hopkins University, Baltimore (USA)
Summary:
The author draws the preliminary conclusion that IQM is a stable dosimetric system and can detect dosimetric deviations caused by small leaf errors in stereotactic radiation therapy. The sensitivity to leaf errors is more pronounced for relatively small fields (i.e., small targets) in SBRT. IQM appears to be more sensitive than Mapcheck2 to small leaf errors in small SBRT targets while showing similar sensitivity in relatively large SBRT targets.
In vivo dosimetry of conventional and rotational IMRT
Author:
Lan Lin, et al, Johns Hopkins University, Baltimore (USA)
Summary:
Evaluation of the accuracy, sensitivity (picket-fence test) and constancy of the IQM for conventional intensity modulated radiation therapy (IMRT) and rotational volumetric modulated arc therapy (VMAT).
Response characteristics of the IQM
Author:
Makan Farrokhkish, Princess Margaret Cancer Centre, Toronto (Canada)
Summary:
Evaluation of the intrinsic and effective spatial sensitivity, response as a function of field size, beam energy and dose rates
Online quality assurance of external beam radiation therapy
Author:
David Hoffman, UC Comprehensive Cancer Center, Sacramento (USA)
Summary:
Evaluation of the stability and accuracy of each IQM system feature.
Initial evaluation of the IQM
Author:
Bruce Perrin, et al, Christie Hospital, Manchester (UK)
Summary:
The authors validated the effectiveness and reliability of the IQM to monitor the correct delivery of daily fractions.
Pre-clinical test of the IQM System
Author:
Stefania Pallotta, AOU Careggi, Florence (Italy)
Summary:
The author states that IQM provides an optimal performance for IMRT QA. Intra-fractional %STD of 0.04% and inter-fractional %STD of up to 0.36% for complex IMRT cases over a period of multiple weeks are the results of this study.
White Paper: IQM Detector characteristics: Signal reproducibility
Author:
iRT Systems, Koblenz (Germany)
Summary:
Investigation of the signal reproducibility of the IQM System
White Paper: IQM Detector characteristics: Signal linearity and dose rate dependence
Author:
iRT Systems, Koblenz (Germany)
Summary:
Investigation of the signal linearity and dose rate dependence of IQM System