Publications

The IQM System has been thoroughly evaluated by some of the world's leading radiation therapy centers.
Select your topic of interest from 15 articles, 16 posters,
18 presentations and 3 technical white papers.

In vivo dosimetry of stereotactic radiation therapy

Author:
Jianguo Qian, et al, Johns Hopkins University, Baltimore (USA)

Summary:
The author draws the preliminary conclusion that IQM is a stable dosimetric system and can detect dosimetric deviations caused by small leaf errors in stereotactic radiation therapy. The sensitivity to leaf errors is more pronounced for relatively small fields (i.e., small targets) in SBRT. IQM appears to be more sensitive than Mapcheck2 to small leaf errors in small SBRT targets while showing similar sensitivity in relatively large SBRT targets.

In vivo dosimetry of conventional and rotational IMRT

Author:
Lan Lin, et al, Johns Hopkins University, Baltimore (USA)

Summary:
Evaluation of the accuracy, sensitivity (picket-fence test) and constancy of the IQM for conventional intensity modulated radiation therapy (IMRT) and rotational volumetric modulated arc therapy (VMAT).

Response characteristics of the IQM

Author:
Makan Farrokhkish, Princess Margaret Cancer Centre, Toronto (Canada)

Summary:
Evaluation of the intrinsic and effective spatial sensitivity, response as a function of field size, beam energy and dose rates

Online quality assurance of external beam radiation therapy

Author:
David Hoffman, UC Comprehensive Cancer Center, Sacramento (USA)

Summary:
Evaluation of the stability and accuracy of each IQM system feature.

Initial evaluation of the IQM

Author:
Bruce Perrin, et al, Christie Hospital, Manchester (UK)

Summary:
The authors validated the effectiveness and reliability of the IQM to monitor the correct delivery of daily fractions.

Pre-clinical test of the IQM System

Author:
Stefania Pallotta, AOU Careggi, Florence (Italy)

Summary:
The author states that IQM provides an optimal performance for IMRT QA. Intra-fractional %STD of 0.04% and inter-fractional %STD of up to 0.36% for complex IMRT cases over a period of multiple weeks are the results of this study.

White Paper: IQM Detector characteristics: Signal reproducibility

Author:
iRT Systems, Koblenz (Germany)

Summary:
Investigation of the signal reproducibility of the IQM System

White Paper: IQM Detector characteristics: Signal linearity and dose rate dependence

Author:
iRT Systems, Koblenz (Germany)

Summary:
Investigation of  the signal linearity and dose rate dependence of IQM System

SIGN UP FOR OUR NEWSLETTER

Sign up for our newsletter. Receive the latest research. Know when we install IQM in your neighbourhood. Get invited to an upcoming event.

Scroll to Top
X